Respiratory Mechanics Module |
METEOR Handheld Monitor |
EZ-FLOW Airway Flow Sensor |
Respiratory Mechanics Evaluation Kit |
|
EZ-FLOW Airway Flow Sensors
|
|
The EZ-Flow flow sensor was
designed to take the cost out of respiratory mechanics monitoring without
compromising performance. The sensor is classified as a fixed orifice,
differential pressure device. It is unique in that it exhibits a
differential pressure / flow curve which is much more linear than other
fixed orifice sensors. This is important in many aspects such as maximizing
low flow resolution and minimizing overall
added airway resistance. |
|
EZ-Flow is available for patients of
all ages and sizes. Our regular EZ-Flow was designed for the
flow/volume ranges of adult and pediatric patients. Our newest EZ-Flow
can handle the most demanding neonate patients, with reliable volume
measurements down to 10cc! We offer our standard tri-tubing for use
with the METEOR Respiratory Mechanics Monitor, or our larger inner diameter
tri-tubing for long term / high humidity environments.

Adult EZ-FLOW SENSOR SPECIFICATIONS
-
Connections: 15 mm ID/22 mm OD patient end by 15 mm
OD.
-
Tapers per ISO 5356-1
-
Dead space: <6.9 ml installed dead space
-
Sensor material: Makrolon polycarbonate
-
Tubing material: 2 meters medical grade
PVC
-
Tubing keyed for proper placement into monitor
-
Sensor Length: 3.25" (6.35 cm)
-
Sensor Weight: 6.3 grams minus tubing
Airway Flow
-
Range: ±2 to 180 LPM (32-3000 ml/s)
-
Total Added Airway resistance: 1 cmH2O added @ 60 LPM (1 LPS)
-
Accuracy: ±5% of reading, or 0.5 LPM (whichever is
greater)
The stated accuracy specifications
apply only to properly placed airway adapter in
patient breathing circuit.
Features
-
Designed to work with any pediatric-adult
ventilator.
-
Simplifies measurement of respiratory mechanics
maneuvers.
-
Intended for continuous monitoring of pediatric or
adult patients with artificial airways and tidal volumes greater than 100
ml.
-
Based on patented flow sensor technology that adds
minimal resistance to the patient’s airflow.
Neonatal EZ-FLOW SENSOR SPECIFICATIONS
-
Connections: 15 mm ID /
OD.
-
Tapers per ISO 5356-1
-
Dead space: <0.75 ml installed dead space
-
Sensor material: Makrolon polycarbonate
-
Tubing material: 2 meters medical grade
PVC
-
Tubing keyed for proper placement into monitor
-
Sensor Length: 2.0" (5.1 cm)
-
Sensor Weight: 4.0 grams minus tubing
Airway Flow
-
Range: ±0.25 to 40 LPM (4-667ml/s)
-
Total Added Airway resistance: 0.8 cmH2O added @ 10 LPM
-
Accuracy: ±5% of reading, or 0.125 LPM (whichever
is greater)
-Sale of Neonate EZ-Flow in the USA pending clearance by the
US Food and Drug Administration
-The stated accuracy specifications
apply only to properly placed airway adapter in patient breathing circuit.
Features
-
Designed to work with neonate ventilators.
-
Simplifies measurement of respiratory mechanics
maneuvers.
-
Intended for continuous monitoring of neonatal
patients with artificial airways and tidal volumes under 100 ml.
-
Based on patented flow sensor technology that adds
minimal resistance to the patient’s airflow.
|
|

Adult Positive Flow Curve |

Adult Negative Flow Curve |

Neonate Positive Flow Curve |

Neonate Negative Flow Curve |